HPLC determination of oxadiazon in commercial pesticide formulations
2008
A simple, fast and precise high performance liquid chromatographic (HPLC) procedure has been developed for the determination of oxadiazon in emulsifiable concentrated pesticide formulations. 20 µL of diluted sample in acetonitrile were injected in a Kromasil C18 (250 ×am injetados em uma coluna Kromasil C18 4.6 mm, 5 µm) column, using acetonitrile:water (80:20) as mobile phase at 1 mL min-1 flow rate and oxadiazon determined by absorbance measurement at 292 nm. A theoretical limit of detection of 0.02 µg mL-1, a limit of quantification of 0.047 µg mL-1, corresponding to a 0.02 and 0.07% m/v in the original sample, and a relative standard deviation of 0.08% for three replicate analysis of sa…